4001 Series

MULTIPARAMETER PHOTOMETRIC ANALYSER FOR CHLORINE AND OTHER DISINFECTANTS

FEATURES

AVAILABLE VERSIONS

APPLICATIONS

TECHNICAL SPECIFICATION

Feature DESCRIPTION
Operating Temperature 0...50°C
Storage and Transport -25...65°C
Humidity 10...95% non-condensing
Power Supply 85...275Vac 50-60Hz
Power Consumption 66 W
Electrical Protection UL6950-1, TUV EN60950, EN 55022 Class B, EN61000, ENV50204, EN55024
Feature DESCRIPTION
Display LCD STN with white backlight
Resolution 240x128 pixels
Languages Italian, English, French, German, Spanish
Keypad 4 bubble-keys [6] [5] [GRAPH/USB] [ESC/MODE] [ENTER/CAL]
Data Logger Internal flash 4Mbit memory equal to 16000 records with a recording interval of 01:00...99:99 min
Recording Method Circular (F.I.F.O.) or filling
Display of Stored Data In tabular and graphic form (1 for each parameter)
Analogue Outputs 1 for each parameter measured (excluding Combined chlorine)
Type 0/4...20 mA galvanically isolated
Programming Limits Lower/upper/reverse
Maximum Load 500 Ohm
Alarm Output According to NAMUR 2.4 mA (with range 4/20mA)
PID Control Activation on the pH output
Set Point Relay Outputs Two for primary measure + two for pH measure (only mod. 4001-3)
Programming Hysteresis, operating time and daily/hourly activation non-subject to the measured value: ON – OFF: 00.00...05.00 ppm Cl2 / 00.00...14.00 pH
Operating Time 000...999 sec.
Max Resistive Load Relay 5A at 230Vac
Alarm Relay Output Cumulative ON-OFF for: Min/Max, set point delay, faults (no water, reagents finished, projector burned, cell dirty)
Delay Time 00:00...59:99 mm:ss with minimum steps of 15 seconds
Auxiliary Relay Output Programmable as: Set point for Temperature measurement or timed activation (programmable frequency and activation time)
Digital Input Clean contact for disabling dosages
RS485 Serial Input Modbus RTU Protocol (1200...38400 Baud rate) for set-up, real-time status or downloading data
Dimensions (LxHxP) 598x601x190 mm
Total Width 598 mm