S401 & S406 & S408

SERIES OF PH & ORP ELECTRODES

The electrodes listed below are all of the combined type (measurement+reference), without maintenance, and are classified by their construction features, which makes them adaptable to multiple applications.

pH MODELS & APPLICATIONS

ORP MODELS & APPLICATIONS

AD SERIES DIGITISER

The AD Series Chemitec digitisers convert the signals of the common pH and ORP electrodes into serial signal with standard Modbus RTU protocol, allowing connection to the 80 and 50 Series plug & play
multiparametric instrument.

TECHNICAL SPECIFICATION

Model MEASURING RANGE OPERATING TEMPERATURE MAXIMUM PRESSURE MIN. LIQUID CONDUCTIVITY
S401 VG 0...14 pH ±2000 mV 0...80°C 6 bar 5 µS/cm
S408 MEC 0...14 pH ±2000 mV 0...130°C 16 bar 50 µS/cm
S401 LC 0...14 pH ±2000 mV 0...100°C 6 bar 0.1 µS/cm
OSM S401 LC 2...14 pH ±2000 mV 0...60°C 16 bar 2 µS/cm
S401 VG HTAJ 0...14 pH ±2000 mV -5...135°C 10 bar 50 µS/cm
S406 VG 0...14 pH ±2000 mV 0...80°C 6 bar 5 µS/cm
S406 POL 0...14 pH ±2000 mV 0...130°C 16 bar 2 µS/cm
S406 OXT 0...14 pH ±2000 mV -5...135°C 16 bar 50 µS/cm
S406 VG HTAJ 0...14 pH ±2000 mV -5...135°C 10 bar 50 µS/cm