DIGITAL UV PHOTOMETRIC SENSORS
New generation of spectral sensors for the online measurement of nitrogen and carbon compounds. Exclusive nanocoated optical lenses and compressed air for extended operating time without cleaning. Optical path length adaptable according to application.
FEATURES
- Ion-selective electrodes
- Stable and sensitive sensors
- All specific electrodes are individually replaceable
- High capacity for pollutant compensation
APPLICATIONS
- Industrial water and chemical industry
- Drinking water
- Wastewater
UV NO3 FEATURES
- Innovative UV photometric technology
- Four detection channels
- Precise optical determination
- Internal temperature correction increases stability of measured values
UV SAC254 FEATURES
- Long life, UV-LED technology
- Robust design
- High energy efficiency
- Automatic turbidity compensation via second measuring channel
- SAC254 measurement correlated displayed with BODeq, CODeq or TOCeq values by means of appropriate internal temperature laboratory checks
TECHNICAL SPECIFICATION
Feature | S480 UV NO3 | S480 UV SAC254 |
---|---|---|
Light Source | Xenon flash lamp | 2 LED technology (254 nm, 530 nm) |
Detector Type | 4 photodiodes & filter | Photodiode |
Measurement Method | Attenuation | Attenuation |
Optical Path | 0.3, 1, 2, 5, 10, 50 mm | 1, 2, 5, 10, 50 mm |
Measurement Parameter | NO3-N, NO3, NOx-N, NOx (Calibrated with standard solution NO3) | SAC254, CODeq, BODeq, TOCeq |
Precision | ± (5 % + 1 mg/L NO3-N) with 1 mm path | 2.5 m⁻¹ SAC254 at 1 mm path |
Turbidity Compensation | Yes | Yes |
Response Time (T100) | 20s | 4s |
Measurement Interval | ≥ 10 s | ≥ 2 s |
Dimensions (L x Ø) mm | 470x48 mm (10 mm path) | 300x48 mm (with 10 mm path) |
Weight | SS ~ 3 kg (titanium 2 kg) | SS ~ 2.7 kg (titanium 1.9 kg) |
Absorption | ≤ 7 W | ≤ 1 W |
MEASURING RANGE S480 UV NO3
Optical Path | Measuring Range (NO3-N) |
---|---|
1 mm | 0.5...60 mg/l NO3-N |
10 mm | 0.05...6 mg/l NO3-N |
MEASURING RANGE S480 UV SAC254
Measuring Range: S480 UV NO3
Optical Path | Measurement Range (NO3-N) |
---|---|
1 mm | 0.5...60 mg/l |
10 mm | 0.05...6 mg/l |
Measuring Range: S480 UV SAC254
Path (mm) | Measurement Parameter | Measurement Range |
---|---|---|
1 | SAC254 | 5...1500/m |
1 | CODeq** | 5...2200 mg/l |
1 | BODeq** | 2.5...700 mg/l |
1 | TOCeq** | 3...880 mg/l |
10 | SAC254 | 0.5...150/m |
10 | CODeq** | 0.8...220 mg/l |
10 | BODeq** | 0.25...70 mg/l |
10 | TOCeq** | 0.3...90 mg/l |
** based on KHP (Note: 100 mg COD-standard-solution corresponds to 85 mg/l KHP)